-In our search, we identified three models of large Ultralow temperature freezer ​-86 C that are above the basic laboratory standards, but not of pharmaceutical grade. ​They were compared ​for characteristics available from manufacturer ​websitesor communicated by the companies ​representative.+In our search, we identified three models of large ultralow (-86 C) temperature freezers ​that are above the basic laboratory standards, but not of pharmaceutical grade. ​We compared ​the characteristics available from the manufacturers' ​websites ​and/or communicated ​to us by the companies' representatives.
  
-^Specifications ​    ^Innova New Brunswick Scientific model U725     ​^Premium New Brunswick Scientific modele U700     ​^Thermo Forma; Fisher FOR990 ​    ​^Thermo Forma; Fisher FOR956 ​    ^+^     ​^Innova New Brunswick Scientific model U725     ​^Premium New Brunswick Scientific modele U700     ​^Thermo Forma; Fisher FOR990 ​    ​^Thermo Forma; Fisher FOR956 ​    ^
 ^Capacity litre     |725 l (25.6 pi cu)     |700 l (25 pi cu)     | 651.3L (23 pi cu)     | 651.3L (23 pi cu)     | ^Capacity litre     |725 l (25.6 pi cu)     |700 l (25 pi cu)     | 651.3L (23 pi cu)     | 651.3L (23 pi cu)     |
-^Insulation ​    ​|"​Ultra-thin (3.15""​ 88mm thick) vacuum insulation panel encapsulated between a metallic polymer film barrier bonded to polyurethane foam" ​    ​|polyurethane foam 130 mm thick     ​|non-CFC foamed-in-place polyurethane insulation in the cabinet; 12.7 cm thick in cabinet; 11.4 cm in the door.     ​|non-CFC foamed-in-place polyurethane insulation in ^the cabinet; 12.7 cm thick in cabinet; 11.4 cm in the door.     |+^Insulation ​    ​|"​Ultra-thin (3.15""​ 88mm thick) vacuum insulation panel encapsulated between a metallic polymer film barrier bonded to polyurethane foam" ​    ​|polyurethane foam 130 mm thick     ​|non-CFC foamed-in-place polyurethane insulation in the cabinet; 12.7 cm thick in cabinet; 11.4 cm in the door.     ​|non-CFC foamed-in-place polyurethane insulation in the cabinet; 12.7 cm thick in cabinet; 11.4 cm in the door.     |
 ^Temperature control ​    ​|Microprocessor Temperature Control System ​    ​|Microprocessor Temperature Control System ​    ​|Patented refrigerated system improves temperature control for more consistent temperature cycling ​    ​|Patented refrigerated system improves temperature control for more consistent temperature cycling ​    | ^Temperature control ​    ​|Microprocessor Temperature Control System ​    ​|Microprocessor Temperature Control System ​    ​|Patented refrigerated system improves temperature control for more consistent temperature cycling ​    ​|Patented refrigerated system improves temperature control for more consistent temperature cycling ​    |
 ^Air filter ​    |Air filter front-mounted easily accessed ​    |Air filter front-mounted easily accessed ​    ​|Convenient hinged air filter grill allows easy access for inspection cleaning and/or changing. ​    ​|Convenient hinged air filter grill allows easy access for inspection cleaning and/or changing. ​     | ^Air filter ​    |Air filter front-mounted easily accessed ​    |Air filter front-mounted easily accessed ​    ​|Convenient hinged air filter grill allows easy access for inspection cleaning and/or changing. ​    ​|Convenient hinged air filter grill allows easy access for inspection cleaning and/or changing. ​     |
-^Access port     |"Two pass-through ports 3/4"" ​20mm diameter" ​    |"Two access ports ; pass-through ports 3/4"" ​20mm diameter ​(pour mettre une sonde externe et un pour mettre le systeme backup ln2 ou co2)" ​    |Unknown ​    ​|Unknown ​    |+^Access port     |Two pass-through ports 3/4 20mm diameter ​    |Two access ports; pass-through ports 3/4 20mm diameter ​    ​|Unknown ​    ​|Unknown ​    |
 ^Battery back up     ​|Battery back-up system activate alarms and displays temperature ​    ​|Battery back-up system activate alarms and displays temperature ​    ​|Battery back-up system is standard ​     |Battery back-up system is standard ​     | ^Battery back up     ​|Battery back-up system activate alarms and displays temperature ​    ​|Battery back-up system activate alarms and displays temperature ​    ​|Battery back-up system is standard ​     |Battery back-up system is standard ​     |
 ^Refrigerant ​    ​|Non-CFC Non-HCFC ​    ​|Non-CFC Non-HCFC ​    ​|non-CFC refrigerants ​    ​|non-CFC refrigerants ​    | ^Refrigerant ​    ​|Non-CFC Non-HCFC ​    ​|Non-CFC Non-HCFC ​    ​|non-CFC refrigerants ​    ​|non-CFC refrigerants ​    |
Line 14: Line 14:
 ^Lock     ​|Removable keyed lock     ​|Removable keyed lock     ​|Locakble ​    ​|Locakble ​    | ^Lock     ​|Removable keyed lock     ​|Removable keyed lock     ​|Locakble ​    ​|Locakble ​    |
 ^Air circulation ​    ​|Requires just one fan.      |Requires just one fan.      |front-to-back air circulation system ​    ​|front-to-back air circulation system ​    | ^Air circulation ​    ​|Requires just one fan.      |Requires just one fan.      |front-to-back air circulation system ​    ​|front-to-back air circulation system ​    |
-^Voltage stabilizer or compensator ​    ​|Internal voltage stabilizer ​    |OPTIONAL ​    |Automatic Voltage Compensator ​    ​|Automatic Voltage Compensator +^Voltage stabilizer or compensator ​    ​|Internal voltage stabilizer ​optional ​    |optional ​    |Automatic Voltage Compensator ​    ​|Automatic Voltage Compensator ​    | 
-compressor ​    |Low-noise heavy duty commercially available ​    ​|Low-noise heavy duty commercially available Embraco ​    ​|Leading compressors 1 HP (horse power) by Copland. ​     |Leading compressors 1 HP (horse power) by Copland. ​     |+^Compressor ​    |Low-noise heavy duty commercially available ​    ​|Low-noise heavy duty commercially available Embraco ​    ​|Leading compressors 1 HP (horse power) by Copland. ​     |Leading compressors 1 HP (horse power) by Copland. ​     |
 ^Exterior ​    |18 gauge steel 1.2 mm thick. Powder coated scratch and rust resistant. ​    |18 gauge steel 1.2 mm thick. Powder coated scratch and rust resistant. ​    ​|Heavy gauge cold-rolled steel cabinets with powder coat paint finish. Uniform exterior surface resists chipping and rust     ​|Heavy gauge cold-rolled steel cabinets with powder coat paint finish. Uniform exterior surface resists chipping and rust     | ^Exterior ​    |18 gauge steel 1.2 mm thick. Powder coated scratch and rust resistant. ​    |18 gauge steel 1.2 mm thick. Powder coated scratch and rust resistant. ​    ​|Heavy gauge cold-rolled steel cabinets with powder coat paint finish. Uniform exterior surface resists chipping and rust     ​|Heavy gauge cold-rolled steel cabinets with powder coat paint finish. Uniform exterior surface resists chipping and rust     |
 ^Outer door     |One outer door      |One outer door     |Two outer doors (double doors model) ​    |One outer door     | ^Outer door     |One outer door      |One outer door     |Two outer doors (double doors model) ​    |One outer door     |
Line 22: Line 22:
 ^Outer door latch     ​|Molded handle with removable lock. One handed operation easy to open. Can be reopened as frequent as desired due to vent     ​|cam-action reverse cam-action for easy one-handed opening ​    |new latch/​handle design ensure a snug door seal     |new latch/​handle design ensure a snug door seal     | ^Outer door latch     ​|Molded handle with removable lock. One handed operation easy to open. Can be reopened as frequent as desired due to vent     ​|cam-action reverse cam-action for easy one-handed opening ​    |new latch/​handle design ensure a snug door seal     |new latch/​handle design ensure a snug door seal     |
 ^Pressure release ​    ​|Heated Vent with spring-assisted plunger to break icing prevents vacuum formation allows to re-open door quickly ​    ​|Heated Vent with spring-assisted plunger to break icing prevents vacuum formation allows to re-open door quickly ​    ​|Vacuum relief port allows pressure to equalize quickly ​    ​|Vacuum relief port allows pressure to equalize quickly ​    | ^Pressure release ​    ​|Heated Vent with spring-assisted plunger to break icing prevents vacuum formation allows to re-open door quickly ​    ​|Heated Vent with spring-assisted plunger to break icing prevents vacuum formation allows to re-open door quickly ​    ​|Vacuum relief port allows pressure to equalize quickly ​    ​|Vacuum relief port allows pressure to equalize quickly ​    |
-^Compartment ​    ​|Three ​compartment ​with two adjustable-height stainless-steel shelves    |Five compartment ​with Four adjustable-height stainless-steel shelves    |"?Four interior ​compartment ​doors complete with sliding latches; ​ three fully adjustable solid +^Compartment ​    ​|Three ​compartments ​with two adjustable-height stainless-steel shelves ​    ​|Five ​compartments ​with four adjustable-height stainless-steel shelves ​    |Four interior ​compartments ​doors complete with sliding latches; three fully adjustable solid stainless steel shelves ​    |Four interior ​compartments ​doors complete with sliding latches; three fully adjustable solid stainless steel shelves ​    |
-stainless steel shelves" ​    |"Four interior ​compartment ​doors complete with sliding latches; ​ three fully adjustable solid +
-stainless steel shelves" ​    |+
 ^Inner doors     ​|Three inner doors insulated and gasketed with high quality seals. Push-oull latch. ​     |Five inner doors insulated and gasketed with high quality seals. Push-oull latch. ​     |four inner doors     |four inner doors     | ^Inner doors     ​|Three inner doors insulated and gasketed with high quality seals. Push-oull latch. ​     |Five inner doors insulated and gasketed with high quality seals. Push-oull latch. ​     |four inner doors     |four inner doors     |
 ^Interior ​    ​|Stainless-steel interior polished 304L     ​|Stainless-steel interior polished 304L     ​|Stainless steel      |Stainless steel      | ^Interior ​    ​|Stainless-steel interior polished 304L     ​|Stainless-steel interior polished 304L     ​|Stainless steel      |Stainless steel      |
 ^Noise level (Decibel) ​    |59 db     |61.8 db     ​|Acoustical mufflers designed into the base and sound absorbing foam reduce noise level 58 db     ​|Acoustical mufflers designed into the base and sound absorbing foam reduce noise level 58 db     | ^Noise level (Decibel) ​    |59 db     |61.8 db     ​|Acoustical mufflers designed into the base and sound absorbing foam reduce noise level 58 db     ​|Acoustical mufflers designed into the base and sound absorbing foam reduce noise level 58 db     |
 ^Casters ​    ​|Heavy-duty lockable casters with leveling feet     ​|Heavy-duty lockable casters with leveling feet     ​|lockable casters ​    ​|lockable casters ​    | ^Casters ​    ​|Heavy-duty lockable casters with leveling feet     ​|Heavy-duty lockable casters with leveling feet     ​|lockable casters ​    ​|lockable casters ​    |
-^"Max vial (2"" ​boxes) capacity" ​    ​|50400 ​    ​|48000 ​    ​|38000 ​    ​|44000 ​    |+^"Max vial (2 inches ​boxes) capacity" ​    ​|50400 ​    ​|48000 ​    ​|38000 ​    ​|44000 ​    |
 ^Racks per freezer ​    ​|18 ​    ​|30 ​    ​|10+10 ​    ​|10+10 ​    | ^Racks per freezer ​    ​|18 ​    ​|30 ​    ​|10+10 ​    ​|10+10 ​    |
 ^Electrical requirement ​    ​|208/​230V 60 Hz     |120V ou 208/230V 60hz     |120V 50/60 Hz 16/20 (P2)     |120V 50/60Hz 1P     | ^Electrical requirement ​    ​|208/​230V 60 Hz     |120V ou 208/230V 60hz     |120V 50/60 Hz 16/20 (P2)     |120V 50/60Hz 1P     |
 ^Exterior size     ​|195mm H X 102.5mm W X 86.7mm D     |1958 mm H X 1125 mm W X 871  mm D     |1976 x 1036 x 940     |1976 x 1036 x 940 ; 5-20P     | ^Exterior size     ​|195mm H X 102.5mm W X 86.7mm D     |1958 mm H X 1125 mm W X 871  mm D     |1976 x 1036 x 940     |1976 x 1036 x 940 ; 5-20P     |
-^Internal size     ​|1355mm H x 865mmW ​ x 615mm D     ​|1355mm H x 865 mmW  x 615mm D     |1308 x 777 x 643     |1308 x 777 x 643 +^Internal size     ​|1355mm H x 865mmW ​ x 615mm D     ​|1355mm H x 865 mmW  x 615mm D     |1308 x 777 x 643     |1308 x 777 x 643     | 
-shipping ​weight ​    |315 kg     |315 kg     ​|408.2 kg     ​|408.2 kg     | +^Shipping ​weight ​    |315 kg     |315 kg     ​|408.2 kg     ​|408.2 kg     | 
-^Heat release ​(dégagement de chaleur) ​    |2559.10 ​btu/hr      | 750 Watts or 2560 Btu/h     ​|Unknown ​    ​|Unknown ​    |+^Heat release ​    ​|2559.10 ​Btu/hr      | 750 Watts or 2560 Btu/h     ​|Unknown ​    ​|Unknown ​    |
 ^Power consumption kilowatts per 24 hours     |16.4 (a t208V) 15.8 (at 220V)     ​|14.9 ​    ​|probably 18     ​|probably 18     | ^Power consumption kilowatts per 24 hours     |16.4 (a t208V) 15.8 (at 220V)     ​|14.9 ​    ​|probably 18     ​|probably 18     |
 ^Pull down time (from 20 C to -80)     ​|Unknown ​    |5.3 hours     ​|Unknown ​    ​|Unknown ​    | ^Pull down time (from 20 C to -80)     ​|Unknown ​    |5.3 hours     ​|Unknown ​    ​|Unknown ​    |
 ^Warm up time (from -80 to -50)     ​|Unknown ​    ​|Unknown ​    ​|approx 200 hours     ​|approx 200 hours     | ^Warm up time (from -80 to -50)     ​|Unknown ​    ​|Unknown ​    ​|approx 200 hours     ​|approx 200 hours     |
-^BTU absorption per hou     |Unknown ​    ​|Unknown ​    |450 par heure     |450 par heure     |+^BTU absorption per hour     |Unknown ​    ​|Unknown ​    |450 par heure     |450 par heure     |
 ^Warranty ​    |5 year on all componants parts and labour ​    |5 year on all componants parts and labour ​    |1 year warranty on all componants parts and labour ​    |1 year warranty on all componants parts and labour ​    | ^Warranty ​    |5 year on all componants parts and labour ​    |5 year on all componants parts and labour ​    |1 year warranty on all componants parts and labour ​    |1 year warranty on all componants parts and labour ​    |
 ^Backup system ​    ​|Optional CO2 and LN2 safety back-up systems ​    ​|Optional CO2 and LN2 safety back-up systems ​    ​|Optional CO2 and LN2 safety back-up systems ​    ​|Optional CO2 and LN2 safety back-up systems ​    | ^Backup system ​    ​|Optional CO2 and LN2 safety back-up systems ​    ​|Optional CO2 and LN2 safety back-up systems ​    ​|Optional CO2 and LN2 safety back-up systems ​    ​|Optional CO2 and LN2 safety back-up systems ​    |
 ^Data logger ​    ​|remote Accense data logger optionnal ​    ​|remote Accense data logger optionnal ​    ​|Optional digital datalogger or Sensaphone ​    ​|Optional digital datalogger or Sensaphone ​    | ^Data logger ​    ​|remote Accense data logger optionnal ​    ​|remote Accense data logger optionnal ​    ​|Optional digital datalogger or Sensaphone ​    ​|Optional digital datalogger or Sensaphone ​    |
 ^Listing and certifications ​    |CE listed cUL listed meets WEEE directives cUL also meets CSA     |CE listed cUL listed meets WEEE directives cUL also meets CSA     |CSA approved; cUL     |CSA approved; cUL     | ^Listing and certifications ​    |CE listed cUL listed meets WEEE directives cUL also meets CSA     |CE listed cUL listed meets WEEE directives cUL also meets CSA     |CSA approved; cUL     |CSA approved; cUL     |
 +
 +Note: 
 +The access ports can be used to insert an external liquid N2 or CO2 backup system. ​