- | In our search, we identified three models of large Ultralow temperature freezer -86 C that are above the basic laboratory standards, but not of pharmaceutical grade. They were compared for characteristics available from manufacturer websites, or communicated by the companies representative. | + | In our search, we identified four models of large ultralow (-86 C) temperature freezers that are above the basic laboratory standards, but not of pharmaceutical grade. We compared the characteristics available from the manufacturers' websites and/or communicated to us by the companies' representatives. |
- | ^Specifications ^Innova New Brunswick Scientific model U725 ^Premium New Brunswick Scientific modele U700 ^Thermo Forma; Fisher FOR990 ^Thermo Forma; Fisher FOR956 ^ | + | ^ ^Innova New Brunswick Scientific model U725 ^Premium New Brunswick Scientific modele U700 ^Thermo Forma; Fisher FOR990 ^Thermo Forma; Fisher FOR956 ^ |
^Capacity litre |725 l (25.6 pi cu) |700 l (25 pi cu) | 651.3L (23 pi cu) | 651.3L (23 pi cu) | | ^Capacity litre |725 l (25.6 pi cu) |700 l (25 pi cu) | 651.3L (23 pi cu) | 651.3L (23 pi cu) | | ||
^Insulation |"Ultra-thin (3.15"" 88mm thick) vacuum insulation panel encapsulated between a metallic polymer film barrier bonded to polyurethane foam" |polyurethane foam 130 mm thick |non-CFC foamed-in-place polyurethane insulation in the cabinet; 12.7 cm thick in cabinet; 11.4 cm in the door. |non-CFC foamed-in-place polyurethane insulation in the cabinet; 12.7 cm thick in cabinet; 11.4 cm in the door. | | ^Insulation |"Ultra-thin (3.15"" 88mm thick) vacuum insulation panel encapsulated between a metallic polymer film barrier bonded to polyurethane foam" |polyurethane foam 130 mm thick |non-CFC foamed-in-place polyurethane insulation in the cabinet; 12.7 cm thick in cabinet; 11.4 cm in the door. |non-CFC foamed-in-place polyurethane insulation in the cabinet; 12.7 cm thick in cabinet; 11.4 cm in the door. | | ||
^Temperature control |Microprocessor Temperature Control System |Microprocessor Temperature Control System |Patented refrigerated system improves temperature control for more consistent temperature cycling |Patented refrigerated system improves temperature control for more consistent temperature cycling | | ^Temperature control |Microprocessor Temperature Control System |Microprocessor Temperature Control System |Patented refrigerated system improves temperature control for more consistent temperature cycling |Patented refrigerated system improves temperature control for more consistent temperature cycling | | ||
^Air filter |Air filter front-mounted easily accessed |Air filter front-mounted easily accessed |Convenient hinged air filter grill allows easy access for inspection cleaning and/or changing. |Convenient hinged air filter grill allows easy access for inspection cleaning and/or changing. | | ^Air filter |Air filter front-mounted easily accessed |Air filter front-mounted easily accessed |Convenient hinged air filter grill allows easy access for inspection cleaning and/or changing. |Convenient hinged air filter grill allows easy access for inspection cleaning and/or changing. | | ||
- | ^Access port |"Two pass-through ports 3/4"" 20mm diameter" |"Two access ports ; pass-through ports 3/4"" 20mm diameter (pour mettre une sonde externe et un pour mettre le systeme backup ln2 ou co2)" |Unknown |Unknown | | + | ^Access port |Two pass-through ports 3/4 20mm diameter |Two access ports; pass-through ports 3/4 20mm diameter |Unknown |Unknown | |
^Battery back up |Battery back-up system activate alarms and displays temperature |Battery back-up system activate alarms and displays temperature |Battery back-up system is standard |Battery back-up system is standard | | ^Battery back up |Battery back-up system activate alarms and displays temperature |Battery back-up system activate alarms and displays temperature |Battery back-up system is standard |Battery back-up system is standard | | ||
^Refrigerant |Non-CFC Non-HCFC |Non-CFC Non-HCFC |non-CFC refrigerants |non-CFC refrigerants | | ^Refrigerant |Non-CFC Non-HCFC |Non-CFC Non-HCFC |non-CFC refrigerants |non-CFC refrigerants | | ||
Line 14: | Line 14: | ||
^Lock |Removable keyed lock |Removable keyed lock |Locakble |Locakble | | ^Lock |Removable keyed lock |Removable keyed lock |Locakble |Locakble | | ||
^Air circulation |Requires just one fan. |Requires just one fan. |front-to-back air circulation system |front-to-back air circulation system | | ^Air circulation |Requires just one fan. |Requires just one fan. |front-to-back air circulation system |front-to-back air circulation system | | ||
- | ^Voltage stabilizer or compensator |Internal voltage stabilizer optional |OPTIONAL |Automatic Voltage Compensator |Automatic Voltage Compensator | | + | ^Voltage stabilizer or compensator |Internal voltage stabilizer optional |optional |Automatic Voltage Compensator |Automatic Voltage Compensator | |
^Compressor |Low-noise heavy duty commercially available |Low-noise heavy duty commercially available Embraco |Leading compressors 1 HP (horse power) by Copland. |Leading compressors 1 HP (horse power) by Copland. | | ^Compressor |Low-noise heavy duty commercially available |Low-noise heavy duty commercially available Embraco |Leading compressors 1 HP (horse power) by Copland. |Leading compressors 1 HP (horse power) by Copland. | | ||
^Exterior |18 gauge steel 1.2 mm thick. Powder coated scratch and rust resistant. |18 gauge steel 1.2 mm thick. Powder coated scratch and rust resistant. |Heavy gauge cold-rolled steel cabinets with powder coat paint finish. Uniform exterior surface resists chipping and rust |Heavy gauge cold-rolled steel cabinets with powder coat paint finish. Uniform exterior surface resists chipping and rust | | ^Exterior |18 gauge steel 1.2 mm thick. Powder coated scratch and rust resistant. |18 gauge steel 1.2 mm thick. Powder coated scratch and rust resistant. |Heavy gauge cold-rolled steel cabinets with powder coat paint finish. Uniform exterior surface resists chipping and rust |Heavy gauge cold-rolled steel cabinets with powder coat paint finish. Uniform exterior surface resists chipping and rust | | ||
Line 22: | Line 22: | ||
^Outer door latch |Molded handle with removable lock. One handed operation easy to open. Can be reopened as frequent as desired due to vent |cam-action reverse cam-action for easy one-handed opening |new latch/handle design ensure a snug door seal |new latch/handle design ensure a snug door seal | | ^Outer door latch |Molded handle with removable lock. One handed operation easy to open. Can be reopened as frequent as desired due to vent |cam-action reverse cam-action for easy one-handed opening |new latch/handle design ensure a snug door seal |new latch/handle design ensure a snug door seal | | ||
^Pressure release |Heated Vent with spring-assisted plunger to break icing prevents vacuum formation allows to re-open door quickly |Heated Vent with spring-assisted plunger to break icing prevents vacuum formation allows to re-open door quickly |Vacuum relief port allows pressure to equalize quickly |Vacuum relief port allows pressure to equalize quickly | | ^Pressure release |Heated Vent with spring-assisted plunger to break icing prevents vacuum formation allows to re-open door quickly |Heated Vent with spring-assisted plunger to break icing prevents vacuum formation allows to re-open door quickly |Vacuum relief port allows pressure to equalize quickly |Vacuum relief port allows pressure to equalize quickly | | ||
- | ^Compartment |Three compartment with two adjustable-height stainless-steel shelves. |Five compartment with four adjustable-height stainless-steel shelves. |Four interior compartment doors complete with sliding latches; three fully adjustable solid | + | ^Compartment |Three compartments with two adjustable-height stainless-steel shelves |Five compartments with four adjustable-height stainless-steel shelves |Four interior compartments doors complete with sliding latches; three fully adjustable solid stainless steel shelves |Four interior compartments doors complete with sliding latches; three fully adjustable solid stainless steel shelves | |
- | stainless steel shelves |Four interior compartment doors complete with sliding latches; three fully adjustable solid | + | |
- | stainless steel shelves | | + | |
^Inner doors |Three inner doors insulated and gasketed with high quality seals. Push-oull latch. |Five inner doors insulated and gasketed with high quality seals. Push-oull latch. |four inner doors |four inner doors | | ^Inner doors |Three inner doors insulated and gasketed with high quality seals. Push-oull latch. |Five inner doors insulated and gasketed with high quality seals. Push-oull latch. |four inner doors |four inner doors | | ||
^Interior |Stainless-steel interior polished 304L |Stainless-steel interior polished 304L |Stainless steel |Stainless steel | | ^Interior |Stainless-steel interior polished 304L |Stainless-steel interior polished 304L |Stainless steel |Stainless steel | | ||
^Noise level (Decibel) |59 db |61.8 db |Acoustical mufflers designed into the base and sound absorbing foam reduce noise level 58 db |Acoustical mufflers designed into the base and sound absorbing foam reduce noise level 58 db | | ^Noise level (Decibel) |59 db |61.8 db |Acoustical mufflers designed into the base and sound absorbing foam reduce noise level 58 db |Acoustical mufflers designed into the base and sound absorbing foam reduce noise level 58 db | | ||
^Casters |Heavy-duty lockable casters with leveling feet |Heavy-duty lockable casters with leveling feet |lockable casters |lockable casters | | ^Casters |Heavy-duty lockable casters with leveling feet |Heavy-duty lockable casters with leveling feet |lockable casters |lockable casters | | ||
- | ^"Max vial (2"" boxes) capacity" |50400 |48000 |38000 |44000 | | + | ^"Max vial (2 inches boxes) capacity" |50400 |48000 |38000 |44000 | |
^Racks per freezer |18 |30 |10+10 |10+10 | | ^Racks per freezer |18 |30 |10+10 |10+10 | | ||
^Electrical requirement |208/230V 60 Hz |120V ou 208/230V 60hz |120V 50/60 Hz 16/20 (P2) |120V 50/60Hz 1P | | ^Electrical requirement |208/230V 60 Hz |120V ou 208/230V 60hz |120V 50/60 Hz 16/20 (P2) |120V 50/60Hz 1P | | ||
^Exterior size |195mm H X 102.5mm W X 86.7mm D |1958 mm H X 1125 mm W X 871 mm D |1976 x 1036 x 940 |1976 x 1036 x 940 ; 5-20P | | ^Exterior size |195mm H X 102.5mm W X 86.7mm D |1958 mm H X 1125 mm W X 871 mm D |1976 x 1036 x 940 |1976 x 1036 x 940 ; 5-20P | | ||
- | ^Internal size |1355mm H x 865mmW x 615mm D |1355mm H x 865 mmW x 615mm D |1308 x 777 x 643 |1308 x 777 x 643 | + | ^Internal size |1355mm H x 865mmW x 615mm D |1355mm H x 865 mmW x 615mm D |1308 x 777 x 643 |1308 x 777 x 643 | |
^Shipping weight |315 kg |315 kg |408.2 kg |408.2 kg | | ^Shipping weight |315 kg |315 kg |408.2 kg |408.2 kg | | ||
- | ^Heat release (dégagement de chaleur) |2559.10 btu/hr | 750 Watts or 2560 Btu/h |Unknown |Unknown | | + | ^Heat release |2559.10 Btu/hr | 750 Watts or 2560 Btu/h |Unknown |Unknown | |
^Power consumption kilowatts per 24 hours |16.4 (a t208V) 15.8 (at 220V) |14.9 |probably 18 |probably 18 | | ^Power consumption kilowatts per 24 hours |16.4 (a t208V) 15.8 (at 220V) |14.9 |probably 18 |probably 18 | | ||
^Pull down time (from 20 C to -80) |Unknown |5.3 hours |Unknown |Unknown | | ^Pull down time (from 20 C to -80) |Unknown |5.3 hours |Unknown |Unknown | | ||
^Warm up time (from -80 to -50) |Unknown |Unknown |approx 200 hours |approx 200 hours | | ^Warm up time (from -80 to -50) |Unknown |Unknown |approx 200 hours |approx 200 hours | | ||
- | ^BTU absorption per hou |Unknown |Unknown |450 par heure |450 par heure | | + | ^BTU absorption per hour |Unknown |Unknown |450 par heure |450 par heure | |
^Warranty |5 year on all componants parts and labour |5 year on all componants parts and labour |1 year warranty on all componants parts and labour |1 year warranty on all componants parts and labour | | ^Warranty |5 year on all componants parts and labour |5 year on all componants parts and labour |1 year warranty on all componants parts and labour |1 year warranty on all componants parts and labour | | ||
^Backup system |Optional CO2 and LN2 safety back-up systems |Optional CO2 and LN2 safety back-up systems |Optional CO2 and LN2 safety back-up systems |Optional CO2 and LN2 safety back-up systems | | ^Backup system |Optional CO2 and LN2 safety back-up systems |Optional CO2 and LN2 safety back-up systems |Optional CO2 and LN2 safety back-up systems |Optional CO2 and LN2 safety back-up systems | | ||
^Data logger |remote Accense data logger optionnal |remote Accense data logger optionnal |Optional digital datalogger or Sensaphone |Optional digital datalogger or Sensaphone | | ^Data logger |remote Accense data logger optionnal |remote Accense data logger optionnal |Optional digital datalogger or Sensaphone |Optional digital datalogger or Sensaphone | | ||
^Listing and certifications |CE listed cUL listed meets WEEE directives cUL also meets CSA |CE listed cUL listed meets WEEE directives cUL also meets CSA |CSA approved; cUL |CSA approved; cUL | | ^Listing and certifications |CE listed cUL listed meets WEEE directives cUL also meets CSA |CE listed cUL listed meets WEEE directives cUL also meets CSA |CSA approved; cUL |CSA approved; cUL | | ||
+ | |||
+ | Note: | ||
+ | The access ports can be used to insert an external liquid N2 or CO2 backup system. |