^ ^Savant DNA120 (Fisher Mandel) ^Vacufuge Complete system (Eppendorf) ^ ^Volume Range/Tube |0.5 to 2.0 mL |0.2 to 25 ml (max height of 27 mm) | ^Maximum Tube Capacity |"36 tubes 2ml (72 when using the 0.6 ml tubes rotor)" |"70 tubes 2ml (F-45-70-11)" | ^Rotor speed |Unknown |1400 rpm (248g) not adjustable | ^Plate rotor max no plates |2 |2 (A-2-VC ) | ^Vacuum Pump Type |ETFE oil-free diaphragm pump |Chemical-resistant PTFE diaphragm pump | ^Displacement |30 L/min at 50 Hz; 36 L/min at 60 Hz |Max 2.1 cubic meter per hour (depends on the mode: aqueous, alcohol or high vapor pressure) | ^Ultimate Vacuum Level Setting |<10 Torr (13 mbar or 1.3 kPa) |20 mBar (achievable ultimate pressure is 9mBar) | ^Refrigerated Trap Capacity |No cold trap |Unrefrigerated 280 mL trap | ^Heat Setting |Ambient, 43°C; or 65°C with over-temperature safety shut-off |Room temperature; 30°C; 45°C; 60°C | ^Heated Cover |Unknown |no heated lid required | ^Vacuum Control Ramp Rate |Unknown |Automatic | ^Timer Setting |5 minutes to 9 hours and 55 minutes and hold |1 min to 9:59h; infinite; in 1 min increment | ^Memory |no programmes |no programmes | ^Chamber |Teflon-coated aluminum |Chemical-resistant stainless steel chamber | ^Chamber Seal |Silicone rubber |High chemical resistivaty seal | ^Fittings |Polypropylene |All steel | ^Pump heads |2-headed pump with ETFE diaphragms |2-headed pump with PTFE diaphragms | ^Cover |Acrylic lid includes safety interlock |Polycarbonate 1 inch | ^Dimensions (W x D x H) |29 x 63 x 33 cm |32 x 53 x 30 cm | ^Additional ventilation space |8 cm all sides |15 cm on all sides | ^Weight (kg) (without rotor) |36 (shipping 45 kg) |31 (shipping ?) | ^Basic Rotor |RD36 |F-45-48-11 | ^Motor |Maintenance-free induction motor |Unknown | ^Applications |Aqueous and alcohol solvents (No ammonium hydroxide) |Aqueous, alcohol solvents and acids (< 1.0N). Another mode is for high vapour pressure | ^Heat Output |3925 BTU/h |1500 BTU/h max | ^Max power consumption |1150 Watts maximum |350 W (with pump) | ^Electrical requirement |115 VAC / 60 Hz/ 10 amps w plugs NEMA 5-15P with an IEC 60320 C13 connector |120V/60Hz 3A (the same model is also available in 230V) | ^Decibel |<65 dBA |<50 dB (A) with pump running |